When measuring capacitance selecting,the right frequency is
important in obtaining the most accurate measurement results.
Generally,a 1 kHz test frequency is used to measure capacitors
that are 0.01
7-1 Selecting Test Frequency
Capacitance
μF or smaller.For capacitors that are 10 μF or larger,a
lower frequency of 120 Hz is used.Following this trend,high test
frequencies are best for testing very low capacitance components.
For large capacitance components,low frequency would be optimal.
For example,if the capacitance of the component is to be in the
mF range,than selecting 100 Hz or 120 Hz for test frequency
would give much better results.The results will also be obvious
because if the same component was tested with 1 kHz or 10 kHz,
the measured readings may look erroneous on the display.
In all cases,it is best to check with the manufacturer’s data
sheet in order to determine the best test frequency to use for
measurement.